D.I.S products for simple and fast residual gas analysis as well as for partial and total pressure measurement.

D.I.S All-In-One Residual Gas Analyzer

951-S7-09-00003
951-S7-09-0003-ALIOGA
951-S7-09-0003 All-In-One Residual Gas Analyzer
  • The D.I.S All-In-One Residual Gas Analyzer is a multipurpose and easy-to-use device for gas analysis and partial/total pressure measurement.
  • Patented ion pulse technology!
  • Total pressure measurement from atmospheric pressure down to 1x10-10 mbar.
  • Simple and fast residual gas analysis in a broad pressure range from 1x10-4 mbar down to 1x10-10 mbar at high sample rates.
  • Helium leaks detection without additional measurement equipment.
  • Realtime vacuum process monitoring with detection of residual gas including heavy hydrocarbons.
  • Easy partial pressure detection of standard air components.
  • User-friendly and intuitive control software makes the operating of the D.I.S All-In-One Gas Analyzer simple and straightforward.

951-S7-09-00003
Figure 1: D.I.S All-In-One Residual Gas Analyzer with exposed vacuum side of the multipurpose device.

Special Features

  • All-In-One residual gas analytics and partial/total pressure measurement device
  • Total pressure measurement range from atmospheric pressure down to 1 x 10-10 mbar
  • Fast and simple residual gas analysis in a broad pressure range from 1x10-4 mbar down to 1x10-10 mbar
  • Mass resolution (R = m/Δm) of 50 
  • Mass range of up to 250 amu
  • All required command & control components and power supplies onboard
  • User-friendly and intuitive control  software
  • Data communication via ethernet connection. Bluetooth or WiFi on demand.
  • No magnetic fields and no RF-fields

OPTIONAL SUPPLEMENTAL DEVICES and features

  • Computer for control software

  • Bluetooth or WiFi connection

  • Higher mass range 

TECHNICAL DATA

total pressure measurement range from atmospheric pressure down to 1 x 10-10 mbar
pressure operating range for gas analysis from 1 x 10-4 mbar down to 1 x 10-10 mbar and better
mounting flange DN 40 CF
data connection Ethernet
power connection included power adapter with connector plug
mass resolution (R = m/Δm)

50

Δm ≤ 1 amu up to 50 amu

Δm ≤ 2 amu up to 100 amu

Δm ≤ 5 amu up to 250 amu

mass range 1 to 250 amu
max. bakeout temperature 120 ºC (on interface flange)
approx. device size (length x width x height)

230 mm x 110 mm x 110 mm - vacuum side

470 mm x 110 mm x 110 mm - total

More technical data on request.

Datasheet:

In preparation